Connect the sterilizer/aerator to a dedicated vent line
to exhaust ethylene oxide to the outside atmosphere or to an emission
control system. The following requirements for venting the Steri-Vac 5XL
gas sterilizer/aerator must be met..
Ensure the vent line is constructed of straight lengths of copper tubing.
Do not extend the length of line beyond 91 .5m (300 ft). The diameter of
the vent line depends on the length of the line. Refer to the table below
for vent sizes. Use a 3/8 in National Pipe Threath(NPT) connection at the
sterilizer. The vent line contains significant amounts of EQ during the
final purge phase. Do not terminate the vent line within 7.6 m (25 ft) of
any possible source of ignition or any opening to the building interior
such as fresh air inlets, unsealed windows or pedestrian traffic areas.

Install a moisture trap if there are more than three
meters (ten feet) of vertical distance in the vent line. Route the vent
line so that moisture drains toward the trap. Avoid sags or loops in the
vent line to prevent moisture buildup at other points in the line. Ensure
that the vent line is gas-tight from the machine to the outside
atmosphere. Use flanged or compression fittings at the sterilizer and the
moisture trap. Braze or solder all the other vent line fittings .Keep all
of the vent line, with the exception of a turned-down extension
terminating on the roof top or exterior wall, inside the building. This is
to prevent moisture from freezing in the line and blocking the vent. (See
the diagram of vent line terminations in Figures
2 and 3.
|