DESCRIPTION AND FEATURES
The Clinix-R general radiographic system consists of a
600 mA single phase x-ray generator, a four-way floating top, horizontal Bucky table,
integrally mounted tube stand with x-ray tube and collimator, and a two size cassette
sensing wall cassette holder. The basic system has several optional components that may be
added to, or substituted for, basic components to tailor a system to your specific needs.
System construction makes it suitable for use in hospitals as a back up radiographic room,
in outpatient departments, clinics, public health agencies, industrial clinics, or
orthopedic clinics.
GENERATOR
The Clinix-R generator consists of a solid state
control with a separate high voltage transformer. The transformer is housed within the
table body in single room installations. This generator is limited to radiographic
capabilities with no provision for fluoroscopic functions.
CONSTRUCTION
The control can be installed flush against the control
booth wall to conserve valuable floor space. The angulated operators panel swings down and
the console front panel is removable for maintenance access from the front of the unit. In
single room installations the high voltage transformer is located within the table body to
conserve space.
LINE VOLTAGE: Nominal 202 or 277 volts, single phase, 50/60 Hertz, with
± 5%
line regulation. A 50 kVa distribution transformer is recommended.
X-RAY TUBES: The basic system powers a single, double focus x-ray tube
for use with two image receptors.
RADIOGRAPHIC RATINGS: 55 kilowatts
Radiographic only
600 mA © 95 kVp
500 mA@ 110 kVp
300 mA @ 125 kVp
200mA @ 28 kVp ( for mammography) |
|